Building a cell that can be trusted
Before any measurement there is a cell to make, and most of the scatter in published cycling data is made at the bench, not at the cycler. A measurement institute that audited academic laboratories found the repeatability of coin-cell results “moderate to low” and the reproducibility between laboratories poor, and traced it to bespoke assembly methods; its guide treats assembly as a process, with a written procedure, training and a record for every cell.5 We follow that guide here for the commonest format, the CR2032 half cell: 20 mm across and 3.2 mm tall, with the electrode under study facing a disc of lithium metal.5
The cell is built in an argon glovebox with oxygen and water below 0.1 ppm, and with a nitrogen filter, because lithium reacts with nitrogen as well; a dry room controls only moisture and is not enough.5 The parts go in as L1.2 shows: the can, then the gasket with its narrower opening down against the can, the positive electrode, the separator, the electrolyte dropped onto the separator once it is in place, the lithium, two spacers (the top one may carry the wave washer) and the cap; the electrode is aligned by eye through the translucent separator.5 Some laboratories build the same stack cap first, spring and spacers down, negative electrode, separator, positive electrode, can; what matters is that every cell of a batch is built the same way.6 The lithium is punched at 15 mm and 200 µm, on a PEEK or POM base and never on PTFE, with a punch kept for lithium alone; only shiny, smooth, grey foil is used, and anything black, white or reddish is rejected.5 The separator is a polypropylene-polyethylene trilayer, never dried above 50 °C, and filter paper is to be avoided.5 The electrolyte, 100 µL from a calibrated pipette, is dispensed in the same amount in every cell, because any excess is squeezed out at the crimp and the amount then differs from cell to cell.5,7 The negative electrode is cut a little larger than the positive, so that a small misalignment does not leave positive material facing nothing; that is also how commercial cells are built.7
Three habits make the difference. Never mix parts from different suppliers, and record part numbers rather than descriptions: cans, caps, spacers and gaskets differ in diameter, thickness and coating, and a mismatch of about half a millimetre between the parts and the crimper has been found in several laboratories, one of which had its crimper retooled.5 Take a break after every four or five cells, because fatigue is a source of error, and spread a large batch over several days.5 And keep a spreadsheet with, for every cell, a unique number, the date and the assembler, the collector and electrode masses and the formulation, the diameters of the electrodes and separator, the electrolyte, the open-circuit voltage as assembled, and notes.5
A cell is a process: the same parts, the same order, the same amounts, and a record of each one.